:LOCK
(:CALibration:PROTected:LOCK)
Purpose To lock out calibration.
Format
:cal:prot:lock
Query :cal:prot:lock?
Response 0 Calibration unlocked
1 Calibration locked
Description The
:LOCK command allows you to lock out calibration after completing
those procedures. Thus,
:LOCK performs the opposite of sending the code
with the
:CODE command. The :LOCK? query returns calibration lock sta-
tus.
NOTE To unlock calibration, send the
:CODE command with the appropriate password.
Purpose To unlock calibration so that you can perform the calibration procedures.
Example
:CAL:PROT:LOCK Lock out calibration
:SAVE
(:CALibration:PROTected:SAVE)
Purpose To save calibration constants in EEROM after the calibration procedure.
Format
:cal:prot:save
Description The :SAVE command stores internally calculated calibration constants
derived during calibration in EEROM. EEROM is non-volatile memory, and
calibration constants will be retained indefinitely once saved. Generally,
:SAVE is sent after all other calibration steps (except for :LOCK).
NOTE Calibration will be only temporary unless the
:SAVE command is sent to permanently
store calibration constants. Calibration data will not be saved if any errors occurred.
Example
:CAL:PROT:SAVE Save calibration constants
Calibration Reference B-7