EasyManuals Logo
Home>Keithley>Measuring Instruments>2182

Keithley 2182 Service Manual

Keithley 2182
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
Troubleshooting 4-15
Table 4-7
Input protection and low-noise preamplifier analog switching states
Function
and Range
Q666 Q667 Q664 Q665 Q672 Q673 Q674
U645
Pin 1, 16
U645
Pin 8
U645
Pin 9
DCV1:
10mV On On On On Off Off Off On Off Off
100mV On On On On Off Off Off Off Off On
1V On On On On Off Off Off Off Off On
10V On On Off On Off Off Off Off Off On
100V Off Off Off Off On On On Off Off On
DCV2: Q661 Q662 Q671
100mV On On On Off Off On
1V On On On Off Off On
10V On On On Off Off On