Status byte EAV (Error Available) bit
Whenever an error is available in the error queue, the EAV (Error Available) bit (bit 2) of the
status byte will be set. Use the *STB? query to obtain the status byte, then test bit 2 to see if it
is set. If the EAV bit is set, an error has occurred, and you can use the appropriate error query to
read the error and at the same time clear the EAV bit in the status byte.
Generating an SRQ on error
To program the instrument to generate an IEEE-488 bus SRQ (Service Request) when an
error occurs, send the following command: *SRE 4. This command will enable SRQ when the
EAV bit is set. You can then read the status byte and error queue as outlined above to check for
errors and to determine the exact nature of the error.
Error number Error message
+430 "Precal selection error"
+431 "Precal offset DAC zero error"
+432 "ACAL Temperature error"
+438 "Date of calibration not set"
+439 "Next date of calibration not set"
+440 "Gain-aperture correction error"
+449 "10 vdc ch2 Low-Q zero error"
+500 "Calibration data invalid"
Table B-5
Calibration errors (cont.)
B-12 Calibration Reference

Table of Contents

Other manuals for Keithley 2182

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2182 and is the answer not in the manual?

Keithley 2182 Specifications

General IconGeneral
BrandKeithley
Model2182
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals