5. Verify temperature measurement accuracy for each of the temperature settings listed in
Table 1-4. For each measurement:
Set the Model 2182 for the appropriate thermocouple type.
Set the temperature calibrator to the correct thermocouple type and temperature
setting.
Verify that the Model 2182 temperature reading is within the limits given in the
table.
Analog output accuracy
Use the following procedure to verify accuracy of the analog output.
NOTE Be sure that DC voltage measurement accuracy is within required limits before
attempting this procedure.
1. With the power off, connect the Model 2182 CHANNEL 1 input jack to the DC voltage
calibrator, and connect the rear panel ANALOG output jack to the digital multimeter, as
shown in Figure 1-4.
Table 1-4
Temperature measurement accuracy limits
Thermocouple
type
Temperature calibrator
setting (°C)
Model 2182 temperature reading limits
(1 Year, 18°C-28°C)
J, K, N, T
J, K, N, T
J, K, N, T
J, K, N
K, N
E
E
E
R, S
R, S, B
R, S, B
-200
0
400
700
1300
-200
0
1000
0
400
1700
-200.2 to –199.8°C
-0.2 to 0.2°C
399.8 to 400.2°C
699.8 to 700.2°C
1299.8 to 1300.2°C
-200.2 to -199.8°C
-0.2 to 0.2°C
999.8 to 1000.2°C
-0.2 to 0.2°C
399.8 to 400.2°C
1699.8 to 1700.2°C
1-10 Performance Verification

Table of Contents

Other manuals for Keithley 2182

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2182 and is the answer not in the manual?

Keithley 2182 Specifications

General IconGeneral
BrandKeithley
Model2182
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals