Figure 1-4
Connections for analog output accuracy
2. Select the Model 2182 DCV1 measurement function, and choose the 10V range.
3. Press SHIFT then AOUT, then use the RANGE and left and right arrow keys to set the
following:
• STATE: ON
• M: +1.0000000
• B: +00.000000
4. Set the calibrator voltage to 0.00000V, then enable the Model 2182 REL mode. Also
enable REL on the digital multimeter.
5. Verify analog output accuracy for each of the voltages listed in Table 1-5. For each test
point:
• Set the DC calibrator voltage to the indicated value.
• Verify that the digital multimeter voltage reading is within the limits given in the
table.
Table 1-5
Analog output accuracy limits
Calibrator voltage Nominal analog output value Analog output voltage limits
2.500000V
5.000000V
7.500000V
10.00000V
0.25V
0.5V
0.75V
1V
0.24875 to 0.25125V
0.4985 to 0.5015V
0.74825 to 0.75175V
0.998 to 1.002V
Model 2000 DMM
2001 MULTIMETER
SHIFT
CH1REM
TALK
LSTN
SRQ
STAT
REL FILT
4W
BUFFER
MATH
REAR
SCAN
TIME
R
STEP CH2 CH3 C H4 CH5 CH6 CH7
CH8
CH9
CH
10
HOLD
TRIG FAST MED SLO
W
AUTO
ERR
!
CHANNEL 1
CHANNEL 2
2107 Low
Thermal Cable
OUTPUT LO
OUTPUT HI
DC Voltage Calibrator
Connect to
ANALOG OUTPUT
on Rear Panel
BNC to Dual
Banana Adapter
Performance Verification 1-11