Calibration considerations
When performing the calibration procedures:
Make sure that the test equipment is properly warmed up and connected to the
appropriate Model 2182 jack.
Always allow the source signal to settle before calibrating each point.
Use only copper-to-copper connections to minimize thermal EMFs.
Make sure that all connections are clean and free of oxidation to avoid thermal EMFs
that could affect calibration accuracy. Clean oxidation from connector terminals with
DeoxIT cleaning solution.
Keep test connections and all equipment away from drafts to avoid thermal drift.
Do not connect test equipment to the Model 2182 through a scanner or other switching
equipment.
If an error occurs during calibration, the Model 2182 will generate an appropriate error
message. See Appendix B for more information.
WARNING The maximum voltage between terminals to chassis ground is 350V peak.
Exceeding this value may cause a breakdown in insulation, creating a shock
hazard.
CAUTION The maximum voltage between CHANNEL 1 HI and LO is 120V. The
maximum voltage between CHANNEL 2 HI and LO is 12V. Exceeding
these voltage values may result in instrument damage.
Calibration cycle
Perform calibration every two years to ensure the unit meets or exceeds its specifications.
Calibration menu
The calibration menu allows you to access the calibration items summarized in Table 2-1. To
access the calibration menu, press SHIFT then CAL. Use the RANGE and left and right arrow
keys to scroll through selections. Use the ENTER key to choose menu items, and use the EXIT
key to cancel them.
Table 2-1
Calibrating menu
Calibration menu item Description
CAL: DATES Display calibration and calibration due dates.
CAL: COUNT Display number of times unit was calibrated.
CAL: RUN Run calibration procedure.
CAL: TEMP Display internal calibration temperature.
Calibration 2-3

Table of Contents

Other manuals for Keithley 2182

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2182 and is the answer not in the manual?

Keithley 2182 Specifications

General IconGeneral
BrandKeithley
Model2182
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals