Precalibration
1. Connect the calibrator to CHANNEL 1 of the Model 2182 input jack using the low-
thermal cable, as shown in Figure 2-3. Wait two minutes to allow for thermal equilibrium
before proceeding. (See Table 2-3 for Model 2107 cable connections.)
2. Set the calibrator to output DC volts, and turn external sense off.
Figure 2-3
Connections for factory DC volts calibration
NOTE Keep drafts away from low-thermal connections to avoid thermal drift, which could
affect calibration accuracy.
3. Perform the precalibration steps summarized in Table 2-7. For each step:
Set the calibrator to the indicated voltage, and make sure the unit is in operate.
Allow the signal voltage to settle.
Send the indicated programming command.
Wait until the Model 2182 completes each step before continuing.
NOTE Keep the calibrator locked to the 11V range.
Table 2-7
Remote precalibration steps
Precalibration step Calibrator voltage Calibration command
10V 10.000000V :CAL:PROT:PCAL:STEP0 10
0V 0.0000000V :CAL:PROT:PCAL:STEP1 0
1V 1.0000000V :CAL:PROT:PCAL:STEP2 1
!
CHANNEL 1
CHANNEL 2
Model 2182
2107 Low
Thermal Cable
OUTPUT LO
Red (CH1)
Green (CH2)
OUTPUT HI
Black (CH1)
White (CH2)
DC Voltage Calibrator
2-14 Calibration

Table of Contents

Other manuals for Keithley 2182

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2182 and is the answer not in the manual?

Keithley 2182 Specifications

General IconGeneral
BrandKeithley
Model2182
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals