EasyManuals Logo
Home>Keithley>Measuring Instruments>2182

Keithley 2182 Service Manual

Keithley 2182
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #94 background imageLoading...
Page #94 background image
Keithley Instruments, Inc.
Cleveland, Ohio 44139
DRN.
CKD.
APP.
DATE
DATE
DATE
PART NUMBER
LTR REVISIONS APP. DATE
FORM 28777A-SBG
BRUNING 40-21 62198-SBG
SPECIFICATIONS
SPEC-2182A
A Released 29844 SZ 6/11/04
HW 6/11/04
SK 6/11/04
RKN 6/08/04
Rev A
Page 2 of 3
2182A Nanovoltmeter Specifications
Temperature (Thermocouples)
12
(DISPLAYED IN °C, °F, OR K. ACCURACY BASED ON ITS-90, EXCLUSIVE OF
THERMOCOUPLE ERRORS.)
TYPE RANGE RESOLUTION
ACCURACY
90 Day/1 Year 23° ±5°C
Relative to Simulated
Reference Junction
J -200 to +760°C 0.001°C
±0.2°C
K -200 to +1372°C 0.001°C
±0.2°C
N -200 to +1300°C 0.001°C
±0.2°C
T -200 to +400°C 0.001°C
±0.2°C
E -200 to +1000°C 0.001°C
±0.2°C
R 0 to +1768°C 0.1°C
±0.2°C
S 0 to +1768°C 0.1°C
±0.2°C
B +350 to +1820°C 0.1°C
±0.2°C
OPERATING CHARACTERISTICS
13,14
60HZ (50HZ) OPERATION
FUNCTION DIGITS READINGS/s PLCs
DCV Channel 1, 7.5 3 (1.2) 5
Channel 2, 7.5
17,19
6 (1.7) 5
Thermocouple 6.5
18,19
18 (5.5) 1
6.5
18,19,20
45 (7.2) 1
5.5
17,19
80 20.9) 0.1
4.5
16,17,19
115 (28.0) 0.01
Channel 1/Channel 2, 7.5 1.5 (1.2) 5
(Ratio), 7.5
17,19
2.3 (1.7) 5
Delta with 24XX, 6.5
18
8.5 (5.5) 1
Scan 6.5
18,20
20 (7.2) 1
5.5
17
30 (20.9) 0.1
Delta with 622X
4.5
17
6.5
41 (28.0)
47 (40)
22
0.01
1
SYSTEM SPEEDS
13,15
RANGE CHANGE TIME
14
: <40 ms (<50 ms).
FUNCTION CHANGE TIME
14
: <45 ms (<55 ms).
AUTORANGE TIME
14
: <60 ms (<70 ms).
ASCII READING TO RS-232 (19.2K Baud): 40/s (40/s).
MAX. INTERNAL TRIGGER RATE
16
: 120/s (120/s).
MAX. EXTERNAL TRIGGER RATE
16
: 120/s (120/s).
MEASUREMENT CHARACTERISTICS
A-D LINEARITY: 0.8ppm of reading + 0.5ppm of range.
FRONT AUTOZERO OFF ERROR: 10mV – 10V: Add ±(8ppm of reading + 500µV) for <10 minutes and ±1°C.
NOTE: Offset voltage error does not apply for Delta Mode.
AUTOZERO OFF ERROR
10mV: Add ±(8ppm of reading + 100nV) for <10 minutes and ±C.
100mV-100V: Add ±(8ppm of reading + 10µV) for <10 minutes and ±C.
NOTE: Offset voltage error does not apply for Delta Mode.
INPUT IMPEDANCE
10mV-10V: >10G, in parallel with <1.5nF. (Front Filter ON)
10mV-10V: >10G, in parallel with <0.5nF. (Front Filter OFF)
100V: 10MΩ±1%
DC INPUT BIAS CURRENT: <60pA @ 23°C, -10V to 5V.
<120pA @ 23°C, 5 V to 10V.
COMMON MODE CURRENT: <50nA p-p at 50Hz or 60Hz.
INPUT PROTECTION: 150V peak to any terminal, 70V peak Channel 1 LO to
Channel 2 LO.
CHANNEL ISOLATION: >10G.
EARTH ISOLATION: 350V peak, >10G and <150pF any terminal to earth. Add
35pF/ft with Model 2107 Low Thermal Input Cable.

Table of Contents

Other manuals for Keithley 2182

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2182 and is the answer not in the manual?

Keithley 2182 Specifications

General IconGeneral
BrandKeithley
Model2182
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals