2-10 Calibration
19. Press EXIT to return to normal display, then select the 2V source range. Repeat steps 2
through 18 for the 2V range.
20. After calibrating the 2V range, repeat the entire procedure for the 20V and 100V ranges
using Table 2-3 as a guide. Be sure to select the appropriate source range with the EDIT
and RANGE keys before calibrating each range.
21. Press EXIT as necessary to return to normal display.
Table 2-3
Front panel voltage calibration
Source
range
1
Source voltage Multimeter voltage reading
2
0.2V +200.00mV
+000.00mV
-200.00mV
+000.00mV
___________ mV
___________ mV
___________ mV
___________ mV
2V
+2.0000V
+0.0000V
-2.0000V
+0.0000V
___________ V
___________ V
___________ V
___________ V
20V
+20.000V
+00.000V
-20.000V
+00.000V
___________ V
___________ V
___________ V
___________ V
100V
+100.00V
+000.00V
-100.00V
+000.00V
___________ V
___________ V
___________ V
___________ V
1. Use EDIT and RANGE keys to select source range.
2. Multimeter reading used in corresponding calibration step. See procedure.
Step 3: Current calibration
Perform the following steps for each current range, using Table 2-4 as a guide.
1. Connect the Model 2430 to the digital multimeter, as shown in Figure 2-2. Select the
multimeter DC current measurement function.