2600S-901-01 Rev. C / January 2008 Return to Section Topics 16-9
Series 2600 System SourceMeter® Instruments Reference Manual Section 16: Calibration
NOTE (For firmware versions earlier then 1.2.0, substitute the actual
calibration password in the smua.cal.unlock command. The
default Model 2601 password is "KI002601"; the default Model 2602
password is "KI002602".
Figure 16-1
Connections for voltage calibration
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
TSP-Link
LINE FUSE
SLOWBLOW
3.15A, 250V
CAT I
SS
LO LO HIHI GGGG
CHANNEL A
!
!
!
U
L
C
US
LISTED
SourceMeter
4ZA4
CAT I
S
S
LO
LO
HI
HI
G
G
G
G
CHANNEL B
!
R
Input HI
Input LO
Model 2602/2612 Channel A
Connections Shown
LO
HI
Digital Multimeter
S LO
S HI
CHANNEL B
SENSE
LO
LO
GUARD
SENSE
HI
HI
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
TSP-Link
LINE FUSE
SLOWBLOW
3.15A, 250V
!
A LOB LO
R
CHANNEL A
SENSE
LO
LO
GUARD
SENSE
HIHI
CHANNEL B
SENSE
LO
LO
GUARD
SENSE
HI
HI
Model 2636 channel A
connections shown
Input HI
Input LO
LO
HI
Digital Multimeter
S LO
S HI
c. Perform each calibration step listed in Table 16-2 (Model 2601/2601) ,or Table 16-3
(Model 2611/2612), or Table 16-4 (Model 2635/2636) as follows:
1) Select the range being calibrated with this command:
smua.source.rangev = range

Table of Contents

Other manuals for Keithley 2601

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2601 and is the answer not in the manual?

Keithley 2601 Specifications

General IconGeneral
BrandKeithley
Model2601
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals