Section 1: Getting Started Series 2600 System SourceMeter® Instruments Reference Manual
•Select FACTORY to restore factory defaults, and press ENTER.
To select power-on setup:
1. Press the MENU > SETUP > POWERON, and then press ENTER
2. Do one of the following:
• Choose FACTORY to load the original defaults
•Select USER NUMBER (1 through 5) to load a user preference.
3. Press ENTER.
4. Press EXIT to return to the main menu.
Remote operation setups
To save and recall user setups:
The setup.save and setup.recall commands are used to save and recall user setups:
setup.save(n)
setup.recall(n)
where:
n = 1, 2, 3, 4 or 5
Save present setup in memory.
Recall saved user setup from memory.
To restore default setups:
The reset commands return the Series 2600 to the original factory defaults:
reset() Restore all factory defaults.
smua.reset() Restore Channel A defaults.
smub.reset() Restore 2602/2612/2636 Channel B defaults.
To select power-on setup:
The setup.poweron command is used to select which setup to return to on power-up:
setup.poweron = n
where:
n = 0 (*RST defaults)
n = 1 to 5 (user setups 1-5)
Select power-on setup.