Series 3700
System Switch/Multimeter User's Manual Section 5:
Basic Digital Multimeter (DMM) Operation
Document Number: 3700S
-900-01 Rev. A / August 2007 5-
33
Range
Internal
DMM
Comm.
(msec)
SLO
Settle
Time
(msec)
SLO
Measurement
Time
(msec)
Internal
DMM
Comm.
1
-10k
0.06
0.5
min
max
0.06
0.0083
250
0.010
240
100k
0.06
1.0
0.0083
250
0.06
0.010
240
1M
0.06
1.0
0.0083
250
0.06
0.010
240
10M
-
100M
0.06
1.0
0.0083
250
0.06
0.010
240
1. For 10Mand 100M, V
measurement
is made on Input HI. Input Sense HI is unused.
2. Default condition for 4-wire is dmm.opendetector=dmm.ON.
3. For dmm.drycircuit=dmm.ON, I
OPENLEAD
is disable, but print(dmm.opendetector) returns 1.0.
4. Additional cable and 3700 card capacitance can increase settle times, resulting additional
measurement uncertainty. Keithley Instruments recommend the use of Teflon or other low-
dielectric absorption wire insulation for these measurements.
Four-wire dry-circuit open lead detection
The Series 3700 dry-circuit resistance measurement circuitry was designed for low power, low
glitch, and low open voltage applications such as GMRR head testing and air bag / squib test
that require low energy resistance sourcing. For this reason, dmm.opendetecor = dmm.ON is
disabled when dmm.drycircuit = dmm.ON. The I
OPENLEAD
current pulse would exceed the dry-circuit
voltage application.

Table of Contents

Other manuals for Keithley 3730

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3730 and is the answer not in the manual?

Keithley 3730 Specifications

General IconGeneral
BrandKeithley
Model3730
CategoryMultimeter
LanguageEnglish

Related product manuals