5: Basic Digital Multimeter (DMM) Operation Series 3700 System Switch/Multimeter
-38 Document Number: 3700S-900-
The thermocouple open detection times are listed in the following table.
Phase
Internal DMM
Comm.
(msec)
Ioff
Source
Settle
(msec)
T/C
Measurement
Time (msec)
Internal
DMM
Comm.
(msec)
1. Default condition is dmm.opendetector=dmm.ON or 1.
2. For dmm.transducer=dmm.TEMP_THERMISTOR, dmm.TEMP_THREERTD, and
dmm.TEMP_FOURRTD, I
OPENLEAD
phase is disabled, but print(dmm.opendetector) returns 1.0.
3. dmm.opendetector is a common ICL command, shared with fourwireohms. To enable or
disable dmm.opendetector for either function, the appropriate function must be selected before
applying the new dmm.opendetector state. For example, to disable thermocouple open
detection, send dmm.func="temperature" then dmm.opendetector=0.
The following figure provides a schematic representation of the Series 3700 open thermocouple
detection.
Figure 5-22: Simplified T/C open lead detection schematic