5: Basic Digital Multimeter (DMM) Operation Series 3700 System Switch/Multimeter
-52 Document Number: 3700S-900-
4. Set threshold resistance:
Turn the wheel to scroll to the "THRESHOLD" menu item (right most menu item) and press the
ENTER key.
Using the wheel, dial in the desired resistance to be used as a threshold (1–1000.
Press the ENTER key to set.
Press the EXIT key to leave the "CONT ATTR MENU."
5. Apply the resistance to be tested. If using a switching module, perform the following steps to
close the appropriate channel.
a) Use the wheel to dial in the channel number.
b) Press the CLOSE key.
6. Press the TRIG key and observe the displayed reading.
7. To measure other switching channels, repeat steps 5 and 6.
NOTE If the measured circuit is below the set threshold level, the instrument will display the
resistance readings. If the measured circuit is above the threshold level, the
instrument will display the message “OPEN."
8. To disable continuity testing, select a different function (for example, DCV).
9. When finished, press the OPENALL key to open all channels.
NOTE Limits and digital outputs cannot be used when testing continuity with the continuity
(CNT) function. If you need to use these operations, use the 2W function to test
continuity.
NOTE Also see the bus command dmm.threshold for more information on threshold
attributes.
NOTE dmm.threshold is a common ICL command. To enable a unique continuity
threshold, first select the function dmm.func = "continuity", then select the
threshold value. The threshold value will be remembered after exiting when returning
to the function (unless reset).