2-26 Measurement Options Model 6517A Getting Started Manual
Internal scanning
The Model 6517A can be used with a scanner card (such as the Model 6521 or 6522)
installed in the option slot of the instrument. This section provides basic information for
scanning internal channels. If the scanner card is not already installed, refer to the scanner
card instruction manual.
NOTES
For complete information on scanning (internal and external), refer to the scan-
ner card instruction manuals.
The scanner card has to be installed in the option slot in order to access the
menus to configure and run an internal scan.
Congure internal scanner
Perform the following steps to configure an internal scan:
1. Press CONFIG and then CARD to display the scanning options (internal or
external).
2. Place the cursor (using the
and
keys) on INTERNAL and press ENTER to dis-
play the menu items for the internal scanner.
3. Use the following menu structure to configure the internal scanner. To select a
menu item, place the cursor (using the
and
keys) on it and press ENTER.
Parameter values are changed using the cursor keys and the RANGE keys. If you
need more information to navigate through the menu structure, see MENUS.
CHANNELS — Selecting this menu item displays the status (on or off) for each
channel. ON indicates that the channel will be included in the scan, while OFF
indicates that the channel will not be included in the scan. To change the status of a
channel(s), place the cursor on that channel and press a RANGE key. When fin-
ished, press ENTER.
SCAN-MODE — Use this menu item to set the scan mode. The VOLTAGE scan
mode is faster and can be used when break-before-make switching is not required.
VSRC-LIMIT — Use this menu item to enable (YES) or disable (NO) the 200V
V-Source limit. The ±200V limit is used to protect the scanner card. Note that a
V-Source limit (0 to ±1000V) may also be set from the Configure V-Source menu
(see V-SOURCE).
SETTLING-TIME — Use this menu item to set a settling time (0 to 999.999 sec-
onds) for each channel. Be sure to press ENTER after changing the time period.
4. Use the EXIT key to back out of the menu structure.

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals