3-16 Remote Operation Model 6517A Getting Started Manual
Test sequences
:TSEQuence TSEQuence subsystem.
:ARM Arms selected test sequence.
:ABORt Stops a test sequence in process.
:TYPE <name> Specify test; DLEakage, CLEakage, CIResistance,
RVCoefficient, SRESistivity, VRESistivity,
SIResistivity, SQSWeep, STSWeep.
:TSOurce <name> Specify trigger source; MANual, IMMediate, BUS,
TLINk, EXTernal, LCLosure.
:TLINe <NRf> Specify TLINK line; 1 to 6.
DLEakage Diode leakage test path:
:STARt <NRf> Specify start voltage; -1000 to 1000.
:STOP <NRf> Specify stop voltage; -1000 to 1000.
:STEP <NRf> Specify step voltage; -1000 to 1000.
:MDELay <NRf> Specify measure delay; 0 to 99999.9 (sec.).
:CLEakage Capacitor leakage test path:
:SVOLtage <NRf> Specify bias voltage (-1000 to 1000).
:STIME <NRf> Specify bias time; 0 to 99999.9 (sec.).
:DTIMe <NRf> Specify discharge time; 0 to 9999.9 (sec.).
:CIResistance Cable insulation resistance test path:
:SVOLtage <NRf> Specify bias voltage; -1000 to 1000.
:STIME <NRf> Specify bias time; 0 to 99999.9 (sec.).
:RVCoefficient Resistor voltage coefficient test path:
:SVOLtage[1] <NRf> Specify source voltage 1; -1000 to 1000.
:MDELay[1] <NRf> Specify measure delay 1; 0 to 99999.0 (sec.).
:SVOLtage2 <NRf> Specify source voltage 2; -1000 to 1000.
:MDELay2 <NRf> Specify measure delay 2; 0 to 99999.9 (sec.).
:SRESistivity Surface resistivity test path:
:PDTime <NRf> Specify pre-discharge time; 0 to 99999.9 (sec.).
:SVOLtage <NRf> Specify bias voltage; -1000 to 1000.
:STIME <NRf> Specify bias time; 0 to 99999.9 (sec.).
:MVOLtage <NRf> Specify measure voltage; -1000 to 1000.
:MTIMe <NRf> Specify measure time; 0 to 99999.9 (sec.).
:DTIMe <NRf> Specify discharge time; 0 to 99999.9 (sec.).
:VRESistivity Volume resistivity test path:
:PDTime <NRf> Specify pre-discharge time; 0 to 99999.9 (sec.).
:SVOLtage <NRf> Specify bias voltage; -1000 to 1000.
:STIME <NRf> Specify bias time; 0 to 99999.9 (sec.).
:MVOLtage <NRf> Specify measure voltage; -1000 to 1000.
:MTIMe <NRf> Specify measure time; 0 to 99999.9 (sec.).
:DTIMe <NRf> Specify discharge time; 0 to 99999.9 (sec.)

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals