3-24 Remote Operation Model 6517A Getting Started Manual
Storing readings in buffer
The reading buffer in the Model 6517A is flexible and capable. It has three controls, which
are found in the TRACe subsystem. There are commands to control:
The size of the buffer (in readings).
TRACe:POINts <NRf>
Whether or not extra data is stored with each reading (e.g., channel number, time-
stamp). Storing extra data reduces the maximum size of the buffer.
TRACe:ELEMents <name> Specify additional reading elements;
TSTamp, HUMidity, CHANnel,
ETEMperature, VSOurce or NONE.
Select buffer control mode.
TRACe:FEED:CONTrol NEVer immediately stop storing readings
TRACe:FEED:CONTrol NEXT start now, stop when buffer is full
TRACe:FEED:CONTrol ALWays start now, never stop
TRACe:FEED:CONTrol PRETrigger start now, stop when pretrigger is satisfied
The following example program sets up the Model 6517A to take 20 readings as fast as it
can into the buffer, then reads the data back after the buffer has filled. The readings will be
stored with the timestamp, etc., but the program reads back only the reading values and
timestamp.

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals