Model 6517A Getting Started Manual Remote Operation 3-11
If you include the measurement function with this command, the instrument will first go to
the specified function and then perform the one-shot measurement and query, as follows:
:MEASure? Currently selected function
:MEASure:VOLTage[:DC]? Volts function
:MEASure:CURRent[:DC]? Amps function
:MEASure:RESistance? Ohms function
:MEASure:CHARge? Coulombs function
SCPI subsystem commands
Detailed information an all SCPI commands is covered in the User’s Manual. In this sec-
tion, only the commands used for basic operation are summarized.
Defaults
:SYSTem SYSTem Subsystem
:PREset Set the 6517A to the system present defaults. Some
of the selected defaults include: volts, continuous
triggers, normal speed (1.0 PLC), and auto range.
:POSetup <name> Select power-on setups: RST, PRESet, or SAV0
through SAV9. Instrument will power up to the
*RST or system present defaults, or one of the
*SAV0 – *SAV9 setups.
Fresh readings
[:SENSe[1]] SENSe Subsystem
:DATA:FRESh? Requests a “fresh” reading. This command will not
request the same reading twice. If a new reading
has already been triggered, this command will wait
until the new reading is available.
:CALCulate[1] CALCulate Subsystem
:DATA:FRESh? Requests a “fresh” calculated reading. When
CALC1 is enabled, sens:data:fres? is the
reading before calculation and
calc:data:fres? is the reading after calcula-
tion. When CALC1 is disabled, the readings are the
same.