Model 6517A Getting Started Manual Remote Operation 3-15
Trigger model
Performing a system preset (:SYSTem:PRESet) configures the advanced trigger model for
typical instrument operation. After a system preset, many advanced operations can be
accomplished using the initiate commands, :ABORt, and the trigger (measure) layer com-
mands.
:INITiate
[:IMMediate] Take the instrument out of idle (ARM indicator
on).
:CONTinuous <b> Enable or disable continuous initiation.
:ABORt Reset trigger system. Instrument stays in idle if con-
tinuous initiation is disabled (off).
:TRIGger Path to configure the advanced trigger layer:
:SOURce <name> Select control source: IMMediate, TIMer,
MANual, BUS, TLINk, EXTernal or HOLD.
:TIMer <n> Set timer interval (sec): 0.001 to 9999999.999.
:DELay <n> Set delay between measurements (sec) 0 to
999999.999.
:SYSTem
:MACRo
:TRIGger Path to configure basic triggering.
[:EXECute] Select basic triggering.
:MODE <name> Specify trigger mode (CONTinuous or ONE-
Shot).
:SOURce <name> Specify source (IMMediate, MANual, BUS,
EXTernal or TIMer).
:TIMer <n> Specify timer interval; 0.001 to 99999.999
(sec.).

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals