3-12 Remote Operation Model 6517A Getting Started Manual
Function configuration
[:SENSe[1]] SENSe Subsystem
:FUNCtion <name> Select measurement function:
<name> = 'VOLTage[:DC]' Volts function
'CURRent[:DC]' Amps function
'RESistance' Ohms function
'CHARge' Coulombs
function
:VOLTage[:DC] Path to configure volts:
:NPLC <n> Set speed: 0.01 to 10 PLC.
:AUTO <b> Enable or disable auto NPLC.
:RANGe Path to configure measurement range:
[:UPPer] <n> Specify expected reading to select range.
:AUTO <n> Enable or disable auto range.
:REFerence <n> Specify reference (rel): -210 to +210.
:STATe <b> Enable or disable reference (rel).
:ACQuire Use input signal as reference (rel).
:DIGits <n> Specify measurement resolution.
:AUTO <b> Enable or disable auto resolution.
AVERage Path to control digital filter:
:TYPE <name> Select filter (NONE, SCALar or ADVanced).
:TCONtrol <name> Select averaging method (REPeat, or
MOVing).
:COUNt <n> Specify filter count (1 to 100).
:ADVanced Path to configure advanced filter:
:NTOLerance <n> Specify noise tolerance (0 to 100%).
[:STATe] <b> Enable or disable digital filter.
:MEDian Path to control median filter:
:RANK < n> Specify rank for filter (1 to 5).
[:STATe] <b> Enable or disable median filter.
:GUARd <b> Enable or disable guard.
:CURRent:DC Path to configure Amps:
(Volts commands) The commands for Volts (except :GUARd)
apply.
:DAMPing <b> Enable or disable damping.