3-14 Remote Operation Model 6517A Getting Started Manual
V-Source
:SOURce SOURce Subsystem
:VOLTage <n> Specify V-Source level (0 to ±1000V).
:RANGe <n> Select range; ≤100 = 100V range, >100 = 1000V
range.
:LIMit Path for voltage limit:
[:AMPLitude] <n> Specify voltage limit; 0 to 1000 (V).
:STATe <b> Enable or disable limit.
:MCONnect <b> Enable or disable V-Source LO to ammeter LO
connection.
:CURRent Path for current limit:
:RLIMit Path to control resistive current limit:
:STATe <b> Enable or disable resistive I-limit.
:LIMit <b> Path to check current compliance:
[STATe]? <b> Query state of current compliance.
Data store (Buffer)
:TRACe TRACe Subsystem
:ELEMents <name> Select reading elements :TSTamp, HUMidity,
CHANnel, ETEMperature, VSOurce, NONE.
:POINts <n> Specify buffer size.
:FEED Path to control buffer:
:CONTrol <name> Select control mode and enable buffer: NEVer,
NEXT, ALWays, or PRETrigger.
:DATA? Read all readings in buffer.
NOTE See User’s Manual to use :PRETrigger.
Open/close channels and scan
:ROUTe ROUTe Subsystem
:CLOSe <list> Close specified channel.
:STATe? Query the closed channel.
:OPEN <list> Open specified channel.
:OPEN:ALL Open all channels.
:SCAN Path to control scanning:
[:INTernal] <list> Specify internal scan list: 2 to 10 chans.
:EXTernal <list> Specify external scan list: 1 to 400 chans.
:SMEThod <name> Specify settling time for internal card; 0 to
99999.9999 (sec.).
:LSELect <name> Enable the specified scan: INTernal,
EXTernal or NONE.