Channel Calibration
- 30 -
MXD70 pH / Redox
Setup and Operating Guide
Channel Calibration
Temperature has an effect on the output from the pH electrode as well, so it is important that the
buffer temperature is entered into the Calibration Manual Temperature Input if manual temperature
compensation is being employed. Alternatively any automatic temperature compensation element
should be placed in the buffer solution with the pH sensor if Auto TC is being used.
Redox Standards
Unlike pH electrodes, the redox electrode’s slope does not change. Nevertheless, incorrect redox
potentials may be occasionally measured and the cause of these errors is usually a contaminated
platinum (Pt) surface, or a contaminated or plugged reference junction. Calibration of an Redox
Electrode is a single point adjustment calibration only.
For calibration and test purposes, standard solutions at various potentials can be purchased from LTH
Electronics:
Part No. 138/175 124mV Redox Solution (500ml)
Part No. 138/176 358mV Redox Solution (500ml)

Table of Contents

Other manuals for LTH Electronics MXD75

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LTH Electronics MXD75 and is the answer not in the manual?

LTH Electronics MXD75 Specifications

General IconGeneral
BrandLTH Electronics
ModelMXD75
CategoryMonitor
LanguageEnglish

Related product manuals