Channel Setup
- 14 -
MXD70 Electrodeless Conductivity
Setup and Operating Guide
Channel Setup
Electrodeless Conductivity Input Channel Setup
The Channels Setup menu contains the basic configurations for the sensor’s input.
The default security access code is 1000
Main Menu
From the front screen press the menu button to
show the main menu options and select Channels.
/ – Select Option
EXIT
– Return to Front Screen
– Enter Option
Select Channel
Select the electrodeless conductivity input channel
you wish to edit.
/ – Select Option
EXIT
– Return to Main Menu
– Enter Option
Mode
Selecting off-line causes any setpoints associated
with this channel to de-energise. Also causes any
current outputs associated to hold their current
value, useful for when commissioning or calibrating
the instrument.
When the unit is placed in an off-line state “off-line”
will appear in the channel messages section on the
front screen.
If a “Cannot Edit Digital Input Has Control” message
appears, then an associated digital input is currently
controlling the on-line / off-line state of the channel.
/ – Select Option
EXIT
– Cancel
– Save Selection

Table of Contents

Other manuals for LTH Electronics MXD75

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LTH Electronics MXD75 and is the answer not in the manual?

LTH Electronics MXD75 Specifications

General IconGeneral
BrandLTH Electronics
ModelMXD75
CategoryMonitor
LanguageEnglish

Related product manuals