Channel Calibration
MXD70 Auxiliary mA Input
Setup and Operating Guide
- 23 -
Channel Calibration
Calibration Reminder
By enabling the calibration reminder the user can
configure a calibration interval, which when expired
will activate an alarm and channel message on the
front screen.
If the interval has expired and the alarm has
activated, then on the completion of a successful
sensor calibration the next cal date will be
automatically incremented by calibration interval.
If the user completes a successful sensor calibration
before the calibration interval has expired, then the
instrument will ask the user whether they wish to
update the next cal due date by the calibration
interval.
/ – Select Option
EXIT
– Cancel
– Save Selection
Calibration Interval
Sets the interval time for the calibration alarm.
The Next Cal Date will update to show the date of the
next calibration alarm.
/ – Increase / Decrease Digit

– Select Next Digit
EXIT
– Cancel
– Save Value
Next Calibration Date
Sets the exact date of the next calibration alarm.
The Calibration Interval will update to show the
number of days to the next calibration date.
/ – Increase / Decrease Digit or Text

– Select Next Item
EXIT
– Cancel
– Save Entry