EasyManuals Logo
Home>LTH Electronics>Monitor>MXD75

LTH Electronics MXD75 User Manual

LTH Electronics MXD75
631 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #450 background imageLoading...
Page #450 background image
Channel Setup
MXD70 Suspended Solids / Turbidity
Setup and Operating Guide
- 11 -
Channel Setup
Mode
Selecting off-line causes any setpoints associated
with this channel to de-energise. Also causes any
current outputs associated to hold their current
value, useful for when commissioning or calibrating
the instrument.
When the unit is placed in an off-line state “off-line”
will appear in the channel messages section on the
front screen.
If a “Cannot Edit Digital Input Has Control” message
appears, then an associated digital input is currently
controlling the on-line / off-line state of the channel.
/ – Select Option
EXIT
– Cancel
– Save Selection
Units
The channel can be setup to display the reading with
the following units: NTU, FTU, mg/l, g/l, ppm, ppt,
EBC, OD and %.
The relationship between these units and the
incoming probe signal is determined by the
linearisation curve data (see page 14) and range
setting (see next item). They provide a qualitative
rather than quantitative representation of the solids
present in the sample for display purposes and
setpoint / current output processing.
Optionally the instrument can be configured to work
only with the raw probe units by setting the units to
PS.
/ – Select Option
EXIT
– Cancel
– Save Selection

Table of Contents

Other manuals for LTH Electronics MXD75

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LTH Electronics MXD75 and is the answer not in the manual?

LTH Electronics MXD75 Specifications

General IconGeneral
BrandLTH Electronics
ModelMXD75
CategoryMonitor
LanguageEnglish

Related product manuals