EasyManuals Logo

LTH Electronics MXD75 User Manual

LTH Electronics MXD75
631 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #332 background imageLoading...
Page #332 background image
Channel Setup
- 16 -
MXD70 Dissolved Oxygen
Setup and Operating Guide
Channel Setup
Sensor Type
Sets the sensor’s input operating mode to either
Polargraphic or Galvanic.
Only available when probe type is set to “Custom”
/ – Select Option
EXIT
– Cancel
– Save Selection
Bias Voltage
For Polargraphic sensors, the polarising Bias Voltage
can be set using this menu.
/ – Increase / Decrease Digit
– Select Next Digit
EXIT
– Cancel
– Save Value
Membrane Correction Factor
The membrane correction factor is specific to
each make of sensor and characterises the type
and thickness of the membrane material in
terms of how its permeability to Oxygen varies
with temperature.
The Correction Factor can be set at using this
menu.
/ – Increase / Decrease Digit
– Select Next Digit
EXIT
– Cancel
– Save Value

Table of Contents

Other manuals for LTH Electronics MXD75

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LTH Electronics MXD75 and is the answer not in the manual?

LTH Electronics MXD75 Specifications

General IconGeneral
BrandLTH Electronics
ModelMXD75
CategoryMonitor
LanguageEnglish

Related product manuals