Discretes
Modbus RS485 Interface
Operating Guide
- 87 -
Discretes
Dissolved Oxygen
575 Channel 1 Digital Input Switch Setup State 0 = Switch Setup Inactive
680 Channel 2 1 = Switch Setup Active
790 Channel 3
576 Channel 1 Sensor simulation State 0 = Sensor Not Simulating
681 Channel 2 1 = Sensor Simulating
791 Channel 3
577 Channel 1 Pressure simulation State 0 = Pressure Not
Simulating
682 Channel 2
792 Channel 3 1 = Pressure Simulating
578 Channel 1 Temperature simulation State 0 = Temperature Not
Simulating
683 Channel 2
793 Channel 3 1 = Temperature
Simulating
579 Channel 1 Sensor Calibration State 0 = Sensor Not Calibrating
684 Channel 2 1 = Sensor Calibrating
794 Channel 3
580 Channel 1 Temperature Calibration State 0 = Temperature Not
Calibrating
685 Channel 2
795 Channel 3 1 = Temperature
Calibrating
Electrodeless Conductivity
595 Channel 1 Input Switch Setup State 0 = Switch Setup Inactive
700 Channel 2 1 = Switch Setup Active
810 Channel 3
596 Channel 1 Sensor simulation State 0 = Sensor Not Simulating
701 Channel 2 1 = Sensor Simulating
811 Channel 3
597 Channel 1 Temperature simulation State 0 = Temperature Not
Simulating
702 Channel 2
812 Channel 3 1 = Temperature
Simulating
598 Channel 1 Sensor Calibration State 0 = Sensor Not Calibrating
703 Channel 2 1 = Sensor Calibrating
813 Channel 3
599 Channel 1 Temperature Calibration State 0 = Temperature Not
Calibrating
704 Channel 2
814 Channel 3 1 = Temperature
Calibrating
600 Channel 1 Ranging State 0 = Sensor Not Ranging
705 Channel 2 1 = Sensor Ranging
815 Channel 3