Battery Maintenance and Replacement Periodic Maintenance
3 - 8 046-006272-00 A7â„¢ Service Manual
FIGURE 3-5 Internal AG module
FIGURE 3-6 Dust Filter
3.7 Battery Maintenance and Replacement
Maintenance is required for the lithium-ion battery supply in the A7. Replace it every 3 years as
follows:
1. Open the rear cover of the anesthesia system.
2. Open the battery box of the anesthesia system.
3. Remove the old battery.
4. Install the new battery.
5. Close the battery box.
6. Close the rear cover.
7. Use only Mindray approved batteries (P/N: 115-018012-00).
3.8 Functional Tests
Refer to Chapter 4.0 Calibration if any values are out of specification.
38