Disassemble the Breathing System Repair and Disassembly
6 - 78 046-006272-00 A7™ Service Manual
6.3.7 Remove the Contact Switch of the L-shaped Handle
1. Refer to “Remove the Absorbent Canister” on page 6-74. and “Remove the CO2 Bypass
Assembly” on page 6-76. to remove the absorber canister assembly and CO2 bypass assembly.
2. Unscrew the two screws to remove the handle and the cam.
FIGURE 6-166
3. Unplug the cable and, unscrew the two screws to remove the contact switch.
FIGURE 6-167