A7â„¢ Service Manual 046-006272-00 6 - 63
Repair and Disassembly Disassemble the Assemblies
6.2.14 Remove the Built-in Anesthesia Module
1. Unscrew the 6 screws to remove the back cover plate of the cart.
FIGURE 6-138
2. Unplug the related tubes and cables from the anesthesia module.
3. Unscrew the four screws to remove the anesthesia module and the metal sheet bracket as well.
FIGURE 6-139
4. Unscrew the four screws to remove the anesthesia module.
FIGURE 6-140