EasyManuals Logo
Home>Mindray>Medical Equipment>A7

Mindray A7 Service Manual

Mindray A7
482 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #424 background imageLoading...
Page #424 background image
Disassemble the Assemblies Repair and Disassembly
6 - 66 046-006272-00 A7™ Service Manual
6.2.16 Remove Anesthesia Module Inlet Pipeline Assembly
1. Open the service door and refer to“Open the Service Door” on page 6-3.
2. Remove the rear panel and refer to“Remove the Rear Cover Plate Assembly” on page 6-29..
3. Unplug the tubes from the anesthesia module and two-way connector.
FIGURE 6-144
4. Unscrew the two screws to remove the anesthesia module inlet pipeline assembly.
FIGURE 6-145
Cut the cable tie used to fix the filter and disconnect the tube from the filter to remove the filter.

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
BrandMindray
ModelA7
CategoryMedical Equipment
LanguageEnglish

Related product manuals