Functional Tests Periodic Maintenance
3 - 28 046-006272-00 A7™ Service Manual
FIGURE 3-32 Diagnostic Tests > Valves Menu
7. Set safety valve to [ON].
8. Set the PEEP valve pressure to 30 cmH2O.
9. Set the Insp Valve Flow to the following values:
3L/min, 10L/min, 20L/min, 30L/min, 60L/min.
Make sure that the deviation between the measured data of the Flow Sensor, Insp Flow Sensor
and Exp Flow Sensor does not exceed 1 L/min or 5% of the measured value of the calibration
device, whichever is greater. Otherwise, refer to “Flow Calibration (Service)” on page 4-8 to
perform flow calibration again.
10. When finished, reinstall the bellows and the water trap.
3.8.8 Check the Pressure Sensor Accuracy
NOTE: Generally, measurement deviations do not easily occur to pressure
sensors. However, in case of replacing the Ventilator Control Board,
Solenoid Valve Assembly, or Drive Gas Assembly, you need to perform
pressure calibration and check the flow sensors accuracy so as to
confirm the effectiveness of calibration.
NOTE: You can use any pressure meter that has an accuracy of at least ±2% for
the accuracy measurement of the pressure sensors.
To check the measurement accuracy of pressure sensors:
1. Perform pneumatic connections as follows:
The following pictures show the four-way device:

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
BrandMindray
ModelA7
CategoryMedical Equipment
LanguageEnglish

Related product manuals