Disassemble the Breathing System Repair and Disassembly
6 - 82 046-006272-00 A7â„¢ Service Manual
6.3.11 Disassemble the Expiratory/Inspiratory Check Valve
Assemblies
1. Turn the check valve cover counterclockwise to remove it.
FIGURE 6-173
2. Pull out the check valve as shown in the following picture.
FIGURE 6-174