Disassemble the Breathing System Repair and Disassembly
6 - 72 046-006272-00 A7â„¢ Service Manual
6.3.2 Remove the Flow Sensor
1. Turn the locking nuts counterclockwise.
FIGURE 6-154
2. Pull out the inspiration and expiration connectors together with their locking nuts. And then
pull out the flow sensors horizontally.
FIGURE 6-155
3. The following pictures show the appearance of inspiratory and expiratory flow sensor
assemblies.