A7™ Service Manual 046-006272-00 6 - 47
Repair and Disassembly Disassemble the Assemblies
6.2.6 Remove the Module Rack Assembly
1. Open the service door and refer to “Open the Service Door” on page 6-3.
1. Disassemble the rear panel and refer to “Remove the Rear Cover Plate Assembly” on page 6-29.
2. Unplug the cable plug connecting the module rack assembly.
FIGURE 6-102
3. Unscrew the four screws to remove the module rack assembly.
FIGURE 6-103
4. Unscrew the five screws to remove the infrared communication board.
FIGURE 6-104