Disassemble the Assemblies Repair and Disassembly
6 - 60 046-006272-00 A7™ Service Manual
3. Unscrew the two screws on the auxiliary gas outlet assembly to remove the assembly.
FIGURE 6-132
4. Unscrew the three screws to remove the backlight board PCBA of auxiliary gas supply flowmeter
.
FIGURE 6-133
6.2.11 Remove the Rotating Block of Breathing Circle
1. Remove the work surface cover plate and refer to “Remove the Work Surface Cover Plate” on
page 6-28.
2. Unplug the breathing system assembly from the rotating block.
3. Unscrew the four screws and remove the rotating block of breathing circle.
FIGURE 6-134