BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 3-3
Nurse call relay performance test
If the user suspects that the nurse call or analog output does
not function properly.
Analog output performance test
Electrical Safety Tests
Electrical safety
tests
Earth impedance
1. Following any repair or replacement of the power module.
2. When the patient monitor is dropped.
3. At least every two years or as required.
Earth leakage test
Patient leakage current
Patient auxiliary current
Other Tests
Power on test
1. When first installed or reinstalled.
2. Following any maintenance or the replacement of any
main unit parts.
Recorder check Following any repair or replacement of the recorder.
Network print test
1. When first installed.
2. Whenever the printer is serviced or replaced.
Device integration check
1. When first installed.
2. Following any repair or replacement of the external device.
Battery check
Function test
1. When first installed.
2. Whenever a battery is replaced.
Performance test
Once every two months or when the battery run time is
reduced significantly.
Mounting check
1. When first installed.
2. At least every two years or as required.
Note: Performance test is not required for the rSO
2
, and ScvO
2
modules, because the rSO
2
, modules
perform self tests, and the ScvO
2
needs to be calibrated prior to use.
3.1.3.1 Preventative Maintenance Procedures
3.1.4 Visual Inspection
Inspect the equipment for obvious signs of damage. The test is passed if the equipment has no obvious
signs of damage. Follow these guidelines when inspecting the equipment:
Carefully inspect the case, display screen, buttons, knobs, and handle for obvious signs of damage.
Inspect the SMR and parameter modules for obvious signs of damage.
Inspect the power cord, bracket and module accessories for obvious signs of damage.
Inspect all external connections for loose connectors, bent pins or frayed cables.
Inspect all connectors on the equipment for loose connectors or bent pins.
Make sure that safety labels and data plates on the equipment are clearly legible.

Table of Contents

Other manuals for Mindray BeneVision N15

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N15 and is the answer not in the manual?

Mindray BeneVision N15 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N15
CategoryMedical Equipment
LanguageEnglish

Related product manuals