EasyManuals Logo
Home>Mindray>Medical Equipment>BeneVision N15

Mindray BeneVision N15 Service Manual

Mindray BeneVision N15
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #51 background imageLoading...
Page #51 background image
BeneVision N17/BeneVision N15/BeneVision N12 Patient Monitor Service Manual 4-9
4.6.9 Recorder Failures
Failure Description Possible Cause Troubleshooting
No printout
Recorder module
disabled
Verify the recorder status LED is lit.
If it is lit, recover its function in "Factory
Maintenance".
Printing paper jam Reinstall the paper roll properly.
Cable defective or
improperly connected
Verify the cable connecting the recorder and
the main control board is properly
connected.
Verify the connecting cables and connectors
are not damaged.
Recorder failure Replace the recorder.
Main control board
failure
Replace the main control board.
Poor printing effect
Printing paper thermal
coating failure
Replace the printing paper.
Thermal head dirty Clean the thermal head.
Recorder failure Replace the recorder.

Table of Contents

Other manuals for Mindray BeneVision N15

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N15 and is the answer not in the manual?

Mindray BeneVision N15 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N15
CategoryMedical Equipment
LanguageEnglish