4 Thermalert GP Series Operator’s Manual
2.2 MOUNTING INSTRUCTIONS
The GP monitor meets NEMA-12 (IEC529, IP54) requirements (front panel
only) for protection of the electronics when mounted properly. It should be
surface mounted using the flanges and holes provided and mounted in such
a manner to allow the free flow of air around the unit. Ambient tempera-
tures should be kept within the range of 0 to 50°C (32 to 120°F).
Note: Mount the monitor as far away as possible from potential sources of electro-
magnetic interference.
Complete the following steps to mount the GP monitor to a panel:
1. Cut a hole 45 mm high by 92 mm wide (1.8 in by 3.63 in) in the panel.
2. Slip the gasket over the rear of the monitor (see Figure 2), and then back
the monitor into the hole from the front side of the panel. (Mounting
bracket and cosmetic frame are available as accessories.)
3. Attach the fixing clamps as shown in Figure 2 to both sides of the monitor.
Make sure the tab is secure in the hole on both sides.
Gasket
Hole
Fixing Clamp
Tab
Fixing Clamp Screw
Rear Cover
Figure 2: Mounting the Monitor
www. .com
information@itm.com1.800.561.8187

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RayTek Thermalert GP series and is the answer not in the manual?

RayTek Thermalert GP series Specifications

General IconGeneral
BrandRayTek
ModelThermalert GP series
CategoryMeasuring Instruments
LanguageEnglish