Thermalert GP Series Operator’s Manual 21
2.4.7 Peak Hold, Valley Hold, and Average
The Peak Hold,Valley Hold, and Average times are factory set at 0 seconds
and are not activated. If you don’t need to use these signal processing func-
tions, no adjustments are necessary.
Note: Only one signal processing setting can be active at a time.
To set and activate, complete the following steps:
1. Press and hold the button and press the button and release them
both. The Peak Hold display appears (Figure 15).
2. Press the button again and the display shows the current Peak Hold
value in seconds.
3. Set the display by using the and buttons. Note that “000” turns
off Peak Hold.
4. Press the button again and the Valley Hold display (Figure 16)
appears.
O
N
L
C
F
Thermalert GP
Figure 15: Peak Hold Display