RIGOL Front Panel Key Reference
2-52 Userās Guide for DSA1000 Series
Table 2-30 Average number for T-Power measurement
Parameter Explanation
Default 10
Range 1 to 1000
N/A
Knob Step 1
Direction Key Step 1
2. Avg Mode
Sets the Average operation to Exp or Repeat, the default is Exp.
ļ¬ In Exp mode, the result is from the exponent average calculation of the
current N (specified by Avg Num) measurements.
ļ¬ In Repeat mode, the result is the arithmetic average of the current N
(specified by Avg Num) measurements.
3. TP Type
ļ¬ Peak
Displays the power value of the maximum signal between the start and stop
lines. The detector in this type is changed to Pos Peak automatically.
ļ¬ Average
Displays the average power of the signals between the start and stop lines.
The detector in this type is changed to āVoltage Avgā automatically.
ļ¬ RMS
Displays the power which is the root-mean-square value of the signals
between the start and stop lines. The detector in this type is changed to
āRMS Avgā automatically.
4. Start Line
Sets the left margin when measuring the T-Power using the time units. The data
calculated under this measurement is from the start to stop line.
You can use the numeric keys, knob or direction keys to modify this parameter.
For more details, please refer to āParameter Inputā.