RIGOL Front Panel Key Reference
2-44 User’s Guide for DSA1000 Series
Key Points:
This parameter only changes the readout of the TG output power, rather than the
actual value of it.
The offset could be either a positive (gain) or a negative (loss).
You can use the numeric keys, knob or direction keys to modify this parameter.
For more details, please refer toParameter Input.
Table 2-25 TG Level Offset
Parameter
Explanation
Default
0 dB
Range
-200 dB to 200 dB
Unit dB
Knob Step 1 dB
Direction Key Step 10 dB
2.3.4.4 Power Sweep
Enables or disables the power sweep function. If enabled, the power from the TG
varies with the sweep rate of the analyzer within the specified range (from start to
stop frequency) and increases gradually starting from the specified TG power (see
TG level”) within specified power range (seePower Range”). If disabled, the TG
output is fixed at the specified power (see TG level”).
Freq (Hz)
Ampl (dBm)
Start Freq Stop Freq
TG Output
TG Level
Power
Range
Figure 2-14 Power sweep

Table of Contents

Other manuals for Rigol DSA1000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DSA1000 Series and is the answer not in the manual?

Rigol DSA1000 Series Specifications

General IconGeneral
BrandRigol
ModelDSA1000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals