RIGOL Specification
6-2 User’s Guide for DSA1000 Series
6.1 Specification*
6.1.1 Frequency
Frequency
Frequency Range DSA1030, DSA1030-TG 9 kHz to 3 GHz
Frequency Resolution 1 Hz
Internal Frequency Reference
Reference Frequency 10 MHz
Aging Rate <3 ppm/year
Temperature Drift
20 to 30
<3 ppm
Frequency Readout Accuracy
Marker Resolution span / (sweep points-1)
Marker Uncertainty ± (frequency indication ×
frequency reference
uncertainty +1% × span
+ 10% × resolution
bandwidth + marker
resolution)
Marker Frequency Counter
Resolution 1 Hz, 10 Hz, 100 Hz, 1
kHz, 10 kHz, 100 kHz
Uncertainty ±(frequency indication ×
frequency reference
uncertainty + counter
resolution)
Note: Frequency Reference Uncertainty= (aging rate × period since adjustment + temperature
drift).
Note*: If there is no special explanation, the specifications listed are those when the track generator
is off.