Measurement Examples RIGOL
User’s Guide for DSA1000 Series
3-25
3.12 Measuring the Harmonic Distortion
DSA1030/DSA1030-TG that has installed the Advanced Measurement Kit option can
measure the harmonic distortion. In the example below, we use a generator (such as
RIGOL DG3101A) to output a sine signal with 100 MHz and -10 dBm as the signal
under measure.
Method 1: (Use marker function to measure)
Procedures:
1. Device attachment
Connect the output terminal of the generator to the RF INPUT 50Ω terminal on
the front panel of the analyzer.
2. Use marker to measure
(1) Reset the analyzer
Press Preset
(2) Set the parameters
Press FREQ
Press Center Freq and enter 200 MHz.
Press Stop Freq and enter 400 MHz.
Press BW
Press RBW and enter 100 kHz.
Now, you can see the fundamental wave, the second harmonic, and third
harmonic of the signal.
(3) Use the Peak and Marker functions to measure
Press Peak
Execute peak searching and activate a marker whose fundamental wave is
marked with frequency of 100 MHz and amplitude of -10 dBm.
Press Marker
Press Delta
Press Peak
Press Next Peak
TheRefof the marker is then placed at the fundamental wave while the
Deltais at the second harmonic. The measured value would be: the
frequency offset is 100 MHz and the amplitude difference is -31.18 dB.
Press Next Peak

Table of Contents

Other manuals for Rigol DSA1000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DSA1000 Series and is the answer not in the manual?

Rigol DSA1000 Series Specifications

General IconGeneral
BrandRigol
ModelDSA1000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals