RIGOL Measurement Examples
3-32 User’s Guide for DSA1000 Series
3.14 Pass/Fail Measurement
DSA1030/DSA1030-TG that has installed the Advanced Measurement Kit option
supports the pass/fail test function. In the example below, we use a generator (such
as RIGOL DG3101A) to output a sine signal with 100 MHz and -10 dBm as the signal
under measure.
Procedures:
1. Device attachment
Connect the output terminal of the generator to the RF INPUT 50Ω terminal on
the front panel of the analyzer.
2. Measure the signal over the Pass/Fail function
(1) Reset the analyzer
Press Preset
(2) Set the parameters
Press FREQ
Press Center Freq and enter 100 MHz.
Press SPAN
Press Span and enter 10 MHz.
Press Meas
Press Meas Fctn
Press the page down key
Press Pass/Fail
(3) Edit the limit line
Press Meas Setup
Press Edit
Press Point and enter 1.
Press X Axis and enter 95 MHz.
Press Amplitude and enter -55 dBm.
Press Connected and select NO.
Add the points listed in the table below.

Table of Contents

Other manuals for Rigol DSA1000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DSA1000 Series and is the answer not in the manual?

Rigol DSA1000 Series Specifications

General IconGeneral
BrandRigol
ModelDSA1000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals