Measurement Examples RIGOL
User’s Guide for DSA1000 Series
3-19
3.9 Measuring the Occupied Bandwidth
DSA1030/DSA1030-TG that has installed the Advanced Measurement Kit option can
measure the occupied bandwidth. In the example below, we use a generator (such as
RIGOL DG3101A) to output a sine signal with 100 MHz and -10 dBm as the signal
under measure.
Procedures:
1. Device attachment
Connect the output terminal of the generator to the RF INPUT 50Ω terminal on
the front panel of the analyzer.
2. Measure the power distribution
(1) Reset the analyzer
Press Preset
(2) Set the parameters
Press FREQ
Press Center Freq and enter 100 MHz.
Press Meas
Press Meas Fctn and enter the lower menu.
Press OBW
(3) Modify the occupied bandwidth settings
Press Meas Setup
Press OBW Span and enter 2 MHz.
Press Power Ratio and enter 99%.
3. Measurement result
The measurement results shown in the lower window are 53.333 kHz (Occupied
bandwidth) and 0 Hz (Transmit frequency error), see figure below.

Table of Contents

Other manuals for Rigol DSA1000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DSA1000 Series and is the answer not in the manual?

Rigol DSA1000 Series Specifications

General IconGeneral
BrandRigol
ModelDSA1000 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals