3. PANO Sensor Calibration
To optimize PANO image quality in the equipment, the operator should perform the Sensor setting &
calibration operation according to the following procedures.
Regular Sensor Calibration check-up is recommended to perform for image optimization against
temperature and humidity variation.
3.1. PANO Sensor Calibration Standard
PANO Sensor Dark Calibration Standard Value
- Dark data file median value: 700 ~ 2300
PANO Sensor Bright Calibration Standard Value
Bright standard value: 5 points with Cu 1 mm filter
Bright Min Max
Median value
Cal Point 1 50 80
Cal Point 2 80 200
Cal Point 3 200 400
Cal Point 4 400 600
Cal Point 5 600 ~
From the achieved PANO Sensor Bright Calibration data file, the 1 Point
Bright Data file and 5 Point Bright data file should meet the standard
value. Other results from different points Bright Datafile can be varied by
characteristics of each sensor.
3.2. Before PANO sensor calibration
1. Remove the Normal Chin Block from the equipment.
2. Disassemble the Tube case (Refer to Chapter 1. Hardware Part Service Guide).
3. Rename the CAL folder (C:\VCaptureSW\Acquisition\PANO\WidePANO\CAL) name for backup.

Table of Contents

Other manuals for Vatech A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech A9 and is the answer not in the manual?

Vatech A9 Specifications

General IconGeneral
BrandVatech
ModelA9
CategoryMedical Equipment
LanguageEnglish

Related product manuals