EasyManuals Logo
Home>Vatech>Medical Equipment>A9

Vatech A9 Technical Manual

Vatech A9
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #191 background imageLoading...
Page #191 background image
Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
190 / 214
COPYRIGHT© 2020 VATECH.
ALL RIGHTS RESERVED
2. vatech A9(PHT-30CSS) Command-List
Command Remark
[SPM_PANO] Command to enter PANOrama mode
[SPM_CEPH] Command to enter CEPH mode
[SPM_ARCT] Command to enter Arch-CT mode
[SPM_RET_] Move to the shooting start position for each mode
(PANOrama, CEPH, CT)
[SPM_ERDY] Exposure switch press-ready command
[SPM_ERDC] Exposure switch press cancel command
[SPM_PCON] Command to let capture S/W run
[SPM_PCOF]
Command that the exposure software has ended (Cancel
switch press-ready cancel function + laser off function)
[SPM_HV___0000] Tube Voltage Setting Command (ex) 0600 = 60kVp
[SPM_HA___0000] Tube Current Setting Command (ex) 0400 = 4mA
[SPM_LON_] Laser on depending on the capture mode
[SPM_LOF_] Turn off all lasers
[SPM_CPON] sensor power on (Operation for each mode)
[SPM_CPOF] sensor power oFF (Operation for each mode)
[SPM_CSON] CT sensor power on (Independent of the current mode)
[SPM_CSOF] CT sensor power off (Independent of the current mode)
[SPM_PSON] PANOrama sensor power on (Independent of the current
mode)
[SPM_PSOF]
PANOrama sensor power off (Independent of the current
mode)
[SPM_HSON] CEPH sensor power on (Independent of the current mode)
[SPM_HSOF] CEPH sensor power off (Independent of the current mode)
[SPM_PVER] Know your motherboard firmware version (must be used
when the device is down. Transmission during operation may
cause equipment malfunction.)
[SPM_IVER] Knowing the Firmware Version of the Inverter Board
[SPM_TVER]
Know the firmware version of the Main Board, Collimator
Board, Column Board, Inverter Board, MP3, Keypad
[SPM_TUTS] Tube feedback test with exposing switch
[SPM_XTST_0000] exposure time test. Input the command
[SPM_XTST_number] used for authentication
(Number is exposure time, 1 second = 0100, 10 seconds =
1000)
[SPM_XON_] Turn on X-ray exposure
[SPM_XOF_] Turn off X-ray exposure
[SPM_TMP?] Tube temperature requirements

Table of Contents