2. vatech A9(PHT-30CSS) Command-List
Command Remark
[SPM_PANO] Command to enter PANOrama mode
[SPM_CEPH] Command to enter CEPH mode
[SPM_ARCT] Command to enter Arch-CT mode
[SPM_RET_] Move to the shooting start position for each mode
(PANOrama, CEPH, CT)
[SPM_ERDY] Exposure switch press-ready command
[SPM_ERDC] Exposure switch press cancel command
[SPM_PCON] Command to let capture S/W run
[SPM_PCOF]
Command that the exposure software has ended (Cancel
switch press-ready cancel function + laser off function)
[SPM_HV___0000] Tube Voltage Setting Command (ex) 0600 = 60kVp
[SPM_HA___0000] Tube Current Setting Command (ex) 0400 = 4mA
[SPM_LON_] Laser on depending on the capture mode
[SPM_LOF_] Turn off all lasers
[SPM_CPON] sensor power on (Operation for each mode)
[SPM_CPOF] sensor power oFF (Operation for each mode)
[SPM_CSON] CT sensor power on (Independent of the current mode)
[SPM_CSOF] CT sensor power off (Independent of the current mode)
[SPM_PSON] PANOrama sensor power on (Independent of the current
mode)
[SPM_PSOF]
PANOrama sensor power off (Independent of the current
mode)
[SPM_HSON] CEPH sensor power on (Independent of the current mode)
[SPM_HSOF] CEPH sensor power off (Independent of the current mode)
[SPM_PVER] Know your motherboard firmware version (must be used
when the device is down. Transmission during operation may
cause equipment malfunction.)
[SPM_IVER] Knowing the Firmware Version of the Inverter Board
[SPM_TVER]
Know the firmware version of the Main Board, Collimator
Board, Column Board, Inverter Board, MP3, Keypad
[SPM_TUTS] Tube feedback test with exposing switch
[SPM_XTST_0000] exposure time test. Input the command
[SPM_XTST_number] used for authentication
(Number is exposure time, 1 second = 0100, 10 seconds =
1000)
[SPM_XON_] Turn on X-ray exposure
[SPM_XOF_] Turn off X-ray exposure
[SPM_TMP?] Tube temperature requirements