Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
168 / 214
COPYRIGHT© 2020 VATECH.
2. PA Mode Start Position Evaluation
2.1. PA Mode starting position standard
Parameter LSL (Lower Specification Limit) USL (Upper Specification Limit)
PA Start Position 1160 1180
2.2. Phantom Specification
CEPH Phantom QUART Phantom
2.3. Test Method
1. Acquire images in High-Resolution PA Mode.
2. The evaluation is conducted in the same manner as in the evaluation method in Section 1.3.
2.4. Test Result
Parameter LSL Measurement value USL Pass/Fail
PA Start Position 1160 1180