3. Procedure for using remote exposure function
• To use the remote exposure function for the first time, approval and
login of the hospital personnel are required.
• Do not use the remote exposure function when exposure a patient.
•
Before using the remote exposure function, check with the hospital
personnel to make sure there are no people around the equipment/X-ray
room, and use this function.
3.1. When exposure the PANO/CEPH in Console Software
1. Run the Console Software.
2. Open the command input window with the Ctrl + S key.
3. Send the [SPM_XRES] command to activate the remote exposure mode.
4. Set the mode (PANO/CEPH) in the Console Software.
5. Click the Confirm button and Ready button in the Console Software.
6. Automatically start the exposure 3 seconds after the Dark acquisition.
3.2. Remote exposure sequence when acquiring the Bright Calibration
1. Open the VAKCAP.exe file of the applicable mode.
2. Move to the Soft Calib Tab .
3. Acquire DARK.
4. Check the DARK is Normal.
5. Send the [SPM_XRES] command to activate remote exposure mode.
6. Send the [SPM_XRET_0500] command to exposure for 5 seconds.
7. Press the Bright button to acquire Brightness for each point.
3.3. TUTS Test
1. Run the Terminal or the VAKCAP.exe file.
2. Send the [SPM_XRES] command to activate the remote exposure mode.
3. Use the [SPM_XRET_xxxx] command to set the desired exposure time.
ex) [SPM_XRET_0900] = 9 seconds
4. Input the [SPM_TUTS] command to exposure X-ray for the set time.
5. To stop X-ray exposure before the set exposure time during X-ray exposure, stop with
the [SPM_XREH] command.
3.4.
XTST Test
1. Run the Terminal or the VAKCAP.exe file.
2. Send the [SPM_XRES] command to activate the remote exposure mode.
3. Set and input the desired exposure time using the [SPM_XTST_xxxx] command, X-ray
exposure proceeds for the set time.